FDA human factors guidance (“Applying Human Factors and Usability Engineering to Medical Devices”) describes the expectations of the US regulatory authority for the human factors process for medical devices. It puts critical tasks, the use-related risk analysis and human factors validation at the center and is decisive for many US approval procedures.
Significance in the US approval process
Unlike an international standard, the FDA human factors guidance is a guidance document of the authority itself. It is not legally binding in the narrower sense and describes the authority's preferred expectations for human factors documentation; an alternative approach is acceptable if it satisfies the applicable statutes and regulations. The Quality Management System Regulation (QMSR, 21 CFR Part 820), however, is binding and requires usability requirements among the design inputs. In practice, the FDA nevertheless measures submissions against exactly these expectations. For FDA marketing submissions (PMA, 510(k), De Novo, HDE), the supplementary guidance on human factors content (May 2026) describes which information is expected, depending on the use-related risk of the device.
The general guidance is supplemented by product-specific guidance for individual device categories, which formulates additional expectations tailored to the respective type of product. For manufacturers this means that the general guidance forms the starting point, but product-specific documents must be reviewed in addition.
Core elements of the guidance
- Description of users, use environment and user interface as a coherent system, comparable to the use specification according to IEC 62366-1.
- Identification of critical tasks on the basis of a task analysis and risk analysis.
- Use-related risk analysis: the use-related risk analysis, which is often documented in tabular form; the supplementary guidance on human factors content (May 2026) shows an example format.
- Formative evaluations for iterative improvement during development.
- Elimination or reduction of use-related hazards: risk control measures implemented before validation.
- Human factors validation: the empirical assessment of the final device with representative users under realistic conditions, covering all critical tasks. It forms part of design validation. The FDA notes that the term “summative testing” is sometimes defined differently and can omit essential components of validation.
- Documentation: Which human factors information a submission should contain has been described since 2026 by the supplementary guidance “Content of Human Factors Information in Medical Device Marketing Submissions”; see HFE/UE report.
This structure largely follows the same logic as the usability engineering process according to IEC 62366-1, an indication that both sets of rules draw on the same professional foundation even if they use different terms.
Differences from IEC 62366-1
The content proximity between the FDA approach and IEC 62366-1 is great but not complete. In several places the FDA guidance is more concrete than the principle-based standard, for example in the definition of critical tasks (categorized by severity of potential harm) and in the number of validation participants. Since the 2026 revision its definitions are expressly informed by IEC 62366-1:2015+AMD1:2020 and ISO 14971, but not identical in every respect.
The handling of results also differs in practice: Use errors observed in validation must be analyzed and assessed explicitly toward the FDA in the HFE/UE report, including a comprehensible justification of why the remaining risk is acceptable. A purely procedural statement that an error occurred does not meet the regulatory expectation. Merely announcing that an observed error will be mitigated through revised instructions for use or additional training is not sufficient either, unless data show that the measure is effective.
Expectations for the validation study
For human factors validation, the guidance places particular emphasis on the composition of the participants: Each relevant user group, derived from the user profile, is treated as a group to be examined in its own right. As a general minimum the guidance names 15 participants; with several distinguishable user groups, at least 15 per group. For certain device types the recommended minimum can be higher. Groups are considered distinct when their characteristics affect the interaction or when they perform different tasks. Participants should represent the intended users and reside in the United States; exceptions need a case-by-case rationale.
Equally important is the handling of observed difficulties: Not only complete failures but also laborious, hesitant or detour-based successful task completions are relevant for the assessment. The guidance expects the cause (perception, comprehension or execution) to be clarified for each such case, for example through structured post-task interviews after the task has been completed. According to the guidance, thinking aloud (think-aloud) during the task is not acceptable in validation because it does not reflect actual use behavior, although it can be useful in formative evaluation.
Many of the topics addressed by the FDA human factors guidance, known use problems, risk analysis and critical tasks, also support meeting the MDR requirements for ergonomic design and the control of use-related risks according to Annex I.
A joint strategy for both markets
For manufacturers seeking approval in both the EU and the US, a consistently managed process that satisfies both sets of rules is worthwhile. The use specification, task and risk analysis and formative evaluations can generally be used jointly; only the documentation format, terminology and the design of the validation study, in particular the group-wise consideration and the final report, have to be tailored to the respective expectation. For the FDA, participants should generally reside in the United States and labeling and training should correspond to the US market; exceptions need a case-by-case rationale. A joint study for both markets therefore needs a sound justification.
This shared foundation not only saves effort but also increases consistency of content: A product developed and tested according to a uniform logic can be justified coherently toward both authorities, instead of maintaining two parallel, possibly contradictory lines of evidence.
Regulatory reference
The guidance was published by the FDA and addresses manufacturers applying human factors processes during development of new medical devices, with a focus on the user interface including packaging, labeling and instructions for use. It also applies when a marketed device is modified to correct use-related design deficiencies. Which human factors information should accompany a submission (PMA, 510(k), De Novo, HDE) is set out in the supplementary guidance on human factors content (May 2026). It is supplemented by product-specific guidance for individual device categories, which can formulate additional expectations.
The guidance was first issued on February 3, 2016 and revised on August 3, 2026 (under the FDA's Level 2 procedures; docket FDA-2011-D-0469). The revision aligned the definitions and Section 9 (Documentation) with the guidance “Content of Human Factors Information in Medical Device Marketing Submissions” (May 29, 2026) and deleted the former Appendix A. Which human factors information a submission should contain has since followed that second guidance, to which the revised guidance itself refers. The appendices are now organized as A: sample sizes, B: analysis of validation results, C: references.
In terms of content, it corresponds to the European usability engineering process according to IEC 62366-1 and is interlinked with risk management according to ISO 14971 via the use-related risk analysis. For other markets, comparable requirements exist with the NMPA usability guidance.
The FDA human factors guidance describes the regulatory expectations for the human factors process for the US market and is closely related in content to IEC 62366-1 but more concretely formulated, especially for critical tasks and the requirements for validation studies.
For manufacturers in both markets, a consistently managed process with a shared foundation is worthwhile; mainly the group-wise consideration, documentation format and the final report need to be adapted.
Frequently asked questions (FAQ)
Is the FDA human factors guidance legally binding?
Not in the narrower sense, because it is a guidance document and not a regulation. In practice, however, it is decisive because the FDA actually measures submissions against these expectations. Which human factors documentation should accompany a submission, depending on the use-related risk of the device, is described in the supplementary guidance on human factors information in submissions (May 2026).
Can a process run according to IEC 62366-1 be used for FDA approval?
Largely yes. The use specification, task and risk analysis and formative evaluations can generally be used jointly. What mainly needs to be adapted are the group-wise consideration in the validation study, the documentation structure and the final HFE/UE report.
What happens if a difficulty but not a complete failure occurs in validation?
That is also relevant for the assessment. The guidance expects the cause of hesitant, laborious or detour-based successful task completions to be clarified (whether a perception, comprehension or execution problem) and this assessment to be documented in the HFE/UE report.
Does a separate study have to be conducted for each user group?
Not necessarily a separate study, but an independent consideration within the validation: Each relevant user group should be represented by at least 15 participants; for certain devices the minimum can be higher.
Do you want to design your human factors process so that it satisfies both the FDA and European requirements? We guide you consistently through both sets of rules.
More about our usability engineeringSources
- IEC 62366-1:2015+AMD1:2020, Medical devices, Part 1: Application of usability engineering to medical devices
- ISO 14971:2019, Medical devices, Application of risk management to medical devices
- Regulation (EU) 2017/745 on medical devices (MDR)
- FDA Guidance: Applying Human Factors and Usability Engineering to Medical Devices (Final Guidance, February 2016, revised August 2026)
- FDA Guidance: Content of Human Factors Information in Medical Device Marketing Submissions (Final Guidance, May 2026)