Abnormal use (German: abnormale Benutzung) denotes, in the sense of usability and human factors standards, a conscious, intentional act or omission that is contrary to the intended use and lies outside further reasonable user interface-related risk control measures by the manufacturer. It therefore differs fundamentally from use errors, which occur unintentionally, and from reasonably foreseeable misuse.
Read more →
ANSI/AAMI HE75 (Human Factors Engineering Design Reference) is the AAMI standard “Human factors engineering, Design of medical devices”. In its human factors guidance, the FDA lists it as a comprehensive reference with principles, handling of the risk of use errors, design elements, and integrated solutions. The current edition dates from 2025 (as of October 2026).
Read more →
Aseptic presentation (German: aseptische Präsentation) is a term from the context of ISO 11607 and denotes the process in which a sterile medical device is removed or provided from its sterile barrier system without compromising the sterility of the product. It begins with the opening of the packaging and ends with the safe, contamination-free transfer or removal of the product.
Read more →
The CE marking and technical documentation form the manufacturer's proof of conformity under the MDR. The technical documentation under Annex II demonstrates the requirements, the EU declaration of conformity confirms them, and the CE marking indicates them on the device.
Read more →
Clinical evaluation and PMCF is the continuous process required by the MDR with which manufacturers use clinical data to demonstrate the safety, performance and clinical benefit of a device (Art. 61, Annex XIV Part A). The PMCF updates this evidence after placing on the market (Part B).
Read more →
Cognitive load (German: Kognitive Belastung) describes the amount of mental demand a task places on a user's working memory. Because working memory is a tightly limited resource, the likelihood of error rises sharply as soon as the demand reaches this limit, which makes it a central starting point for designing safety-critical user interfaces.
Read more →
The cognitive walkthrough is an analytical evaluation method in which experts step through a concrete user task one step at a time and check, for each step, whether a user would recognize the correct course of action, understand it, and receive feedback that it was successful. The focus is on learnability for users without prior experience.
Read more →
A combination product consists, under 21 CFR 3.2(e), of at least two differently regulated constituent parts, that is, a drug, a device, or a biological product. The FDA uses the primary mode of action to determine the lead center. Human factors is covered by a dedicated FDA guidance.
Read more →
The context of use (German: Nutzungskontext) describes the users, goals, tasks, resources, and environmental conditions under which a product is used. Context analysis (German: Kontextanalyse) refers to the systematic collection and analysis of these conditions.
Read more →
The contextual interview (German: kontextuelles Interview) is a user research method in which users are interviewed and observed at the same time at their real workplace. It reveals how a product is actually used instead of how its use is described, and thus provides the context of use that IEC 62366-1 requires as the basis of every use-related risk analysis.
Read more →
A critical task is, according to the FDA human factors guidance, a user task whose erroneous or omitted performance would or could lead to serious harm to the patient or user. Harm here also includes compromised medical care. The classification remains in place even if alarms, plausibility checks or other risk control measures are present. These reduce the risk but do not remove the criticality of the task itself. Critical tasks determine which use scenarios must be tested in the summative evaluation.
Read more →
The De Novo Classification Request (German: De-Novo-Verfahren) is an FDA regulatory pathway for novel medical devices for which there is no predicate device. The manufacturer requests classification into Class I or II instead of automatic Class III. If the FDA grants it, a new classification regulation is created, with special controls for Class II.
Read more →
Design inputs are the documented requirements for a medical device that serve as the basis for its development. They result, among other things, from the intended purpose (intended use), user needs, regulatory requirements, technical constraints, and findings from risk management and usability engineering. Design outputs are the documented development results that emerge from these requirements. Examples include drawings, specifications, software artifacts, labeling, or instructions for use.
Read more →
Emergency use (German: Notfallnutzung) denotes the use of a medical device in a time-critical situation in which immediate or particularly fast action may be required to prevent or limit serious health consequences. Emergency use is not a defined term in IEC 62366-1 but describes a particular context of use within usability engineering.
Read more →
Ergonomics is, according to the International Ergonomics Association (IEA), the scientific discipline concerned with understanding the interactions among humans and other elements of a system, and at the same time the profession that applies theory, principles, data, and methods to design in order to optimize human well-being and overall system performance.
Read more →
Describing a product as ergonomic is an advertising claim with no defined test basis. Unlike health-related claims about foods, there is no set of rules that defines what evidence must exist before the term may be used.
Read more →
Ergonomics standards define how products and equipment are to be fitted to human capabilities and body dimensions. The most important families are ISO 9241 for human-system interaction, EN 614 for the ergonomic design of machinery, and DIN 33402 (a German national standard) for human body dimensions.
Read more →
Error tolerance (German: Fehlertoleranz) describes the property of a user interface to prevent use errors, catch them immediately, or limit their consequences. It is among the most effective risk reduction measures in usability engineering because it works independently of the user's care and attention, unlike warnings or training.
Read more →
FDA device classification assigns every medical device to one of three risk-based classes under Section 513 of the FD&C Act. The class determines which controls apply and which submission the FDA requires, such as a 510(k) or Premarket Approval.
Read more →
FDA human factors guidance (“Applying Human Factors and Usability Engineering to Medical Devices”) describes the expectations of the US regulatory authority for the human factors process for medical devices. It puts critical tasks, the use-related risk analysis and human factors validation at the center and is decisive for many US approval procedures.
Read more →
A formative usability evaluation (formative evaluation) is a development-accompanying assessment of the user interface or the user-product interaction with the aim of identifying strengths, weaknesses and unforeseen use errors early. It does not serve as the final proof of the safe usability of the user interface but as the basis for its iterative improvement. It is a central component of the usability engineering process according to IEC 62366-1.
Read more →
Grip force and operating forces describe the force a user can exert and the force a product demands of that user. If the required force is above what the weakest intended users can exert, a use-related risk arises, regardless of how comprehensible the operation is otherwise.
Read more →
A hazard-related use scenario is a use scenario in which one or more user actions, in combination with context of use, user characteristics or system properties, can contribute to a hazardous situation or to harm. It describes not only a user action but the complete chain of events consisting of tasks, possible use errors, contributing context factors, hazard, hazardous situation and possible harm. According to IEC 62366-1, they are derived from the identified hazards and hazardous situations and form a central link between usability engineering and risk management according to ISO 14971.
Read more →
A hazardous situation (German: Gefährdungssituation) is, in the risk management of medical devices, the circumstance in which people, property, or the environment are exposed to one or more hazards. It is the link between an existing hazard and a possible harm, and it becomes particularly relevant in usability engineering when it is triggered by user interaction, use errors, or deficiencies in the user interface.
Read more →
The Health Canada Medical Device Licence is the licence that manufacturers need under the Medical Devices Regulations (SOR/98-282) for Class II, III, and IV medical devices. A prerequisite is a quality management system certificate, which must come from an MDSAP-recognized auditing organization. Class I devices do not need a product licence.
Read more →
The HFE/UE report (human factors report) is the summary report expected by the FDA that documents the entire human factors process for a product, from the use specification to the assessment of remaining use-related risks. It is the central evidence document in many US marketing applications.
Read more →
Human factors engineering (HFE) is the systematic application of knowledge about human capabilities, limitations and behavior to the design of products, systems and work environments. In medical technology, HFE pursues the same goal as the usability engineering process according to IEC 62366-1: to enable safe and effective use and to reduce use errors to an acceptable level.
Read more →
Human factors validation testing is the test at the end of product development described in the FDA human factors guidance. It is intended to show that the intended users can use a medical device under the expected conditions of use without serious use errors. The FDA classifies it as part of design validation.
Read more →
The human-machine interface (HMI; German: Mensch-Maschine-Schnittstelle) is the set of all elements through which information is exchanged between user and device: displays, controls, signals, and feedback. It covers both directions of the exchange and is directly safety-relevant in medical technology.
Read more →
IEC 60601-1-6 (Usability Collateral Standard) is the collateral standard to the general standard IEC 60601-1. It sets out a process by which manufacturers analyze, specify, design, verify, and validate the usability of medical electrical equipment, insofar as it relates to basic safety and essential performance.
Read more →
IEC 62366-1 is the internationally authoritative standard for applying usability engineering to medical devices. It describes a structured process that manufacturers use to systematically identify, control and demonstrate use-related risks. Its focus is explicitly on safety, not on comfort or aesthetics. Current version: IEC 62366-1:2015 + AMD1:2020.
Read more →
IEC 62366-2 (officially IEC/TR 62366-2) is a technical report that supplements the normative requirements of IEC 62366-1 with practice-oriented guidance, methods and examples. Unlike Part 1, Part 2 is not mandatory. It explains how the usability engineering process can be carried out in practice without setting normative requirements itself.
Read more →
The indications for use describe, under 21 CFR 814.20(b)(3)(i), the disease or condition that a device diagnoses, treats, prevents, cures, or mitigates, including the patient population. At the FDA they are part of the broader intended use and are stated, for example, on the form FDA 3881.
Read more →
The instructions for use (IFU; German: Gebrauchsanweisung) are part of the user interface and provide the information required for the safe and intended use of a medical device. They are themselves subject to usability assessment: Instructions that are hard to understand cause use errors instead of preventing them.
Read more →
The intended use, called intended purpose in the EU MDR and IVDR (German: Zweckbestimmung, also Verwendungszweck or bestimmungsgemäße Verwendung), describes the medical purpose specified by the manufacturer for which a medical device is intended. It results from the manufacturer's information on the label, in the instructions for use, in promotional and sales materials, and in the clinical evaluation, and it forms the regulatory and technical starting point for classification, risk management, clinical evaluation, and usability engineering.
Read more →
The intended user (German: vorgesehener Benutzer) is the person or group of people whom the manufacturer designates to interact with the medical device, for example as operating professionals, as patients in self-use, or as technical personnel for maintenance, installation, or cleaning. A product can have several intended user groups. The delimitation is important: The intended user describes who is meant to use the product; which characteristics this person brings is the subject of the user profile.
Read more →
ISO 13485 (Quality Management System for Medical Devices) is the international standard with requirements for the quality management system of medical device manufacturers for regulatory purposes. The 2016 edition is authoritative. FDA has incorporated it into the QMSR since February 2, 2026.
Read more →
ISO 20417 (Information to Be Supplied by the Manufacturer) is the international standard for the information a manufacturer supplies with a medical device or accessory: labels, packaging particulars, and accompanying information such as instructions for use. It governs content, not how it is provided. The current edition is the second, from 2026 (as of October 2026).
Read more →
IVDR (Regulation on in vitro diagnostic medical devices, Regulation (EU) 2017/746) is the European regulation for in vitro diagnostic medical devices (IVDs) and governs their development, conformity assessment, placing on the market, market surveillance and performance evaluation within the EU. It replaces the former IVD directive and follows a risk-based regulatory approach with considerably higher requirements for clinical and performance-related evidence, technical documentation and post-market surveillance.
Read more →
Known usability problems (German: bekannte Gebrauchstauglichkeitsprobleme, also known as known use problems) are documented problems, use errors, use difficulties, or use-related hazards that are already known from predecessor products, similar products, vigilance data, market surveillance, literature, complaints, service reports, or earlier usability studies. In the usability engineering process, they serve as a systematic source of information to identify potential use errors early and to incorporate them into the design of the user interface and the use-related risk analysis.
Read more →
A known use error (German: bekannter Anwendungsfehler) denotes an error already observed or documented in the use of a medical device that has occurred with the manufacturer's own medical device, a predecessor product, or a sufficiently comparable product. It can be part of a foreseeable sequence of events and contribute to a hazardous situation or harm, but does not necessarily have to.
Read more →
Label legibility and readability testing asks whether intended users can perceive, find, distinguish and understand information on labels and packaging under real conditions. The formative readability test examines this by observation during development, using information tasks and controlled light variation, and does not replace a summative evaluation.
Read more →
Labeling (German: Kennzeichnung) comprises all information provided to a user on or together with a medical device: product labels, packaging markings, symbols, warnings, and the instructions for use. From a human factors engineering perspective, labeling is part of the user interface and is therefore itself subject to usability assessment, not merely a mandatory regulatory document.
Read more →
The MAUDE database (Manufacturer and User Facility Device Experience) is the publicly accessible US database of the FDA for reports on incidents involving medical devices. In human factors and usability engineering, MAUDE serves as an important signal source for identifying known use-related problems, recurring reported event patterns and possible use-related risks.
Read more →
MDR (Medical Device Regulation, Regulation (EU) 2017/745) is the EU regulation on medical devices. It defines the regulatory requirements for placing on the market, putting into service and market surveillance of medical devices in the European Union, with a strong focus on safety, performance, risk management and usability.
Read more →
The MDSAP (Medical Device Single Audit Program) is a program developed by the IMDRF (International Medical Device Regulators Forum) in which a recognized auditing organization examines a manufacturer's quality management system in a single audit. The audit covers the requirements of all participating regulatory authorities. It does not replace marketing authorization.
Read more →
NMPA usability guidance is the guideline of the Chinese regulatory authority National Medical Products Administration (NMPA), issued by its evaluation center CMDE, for assessing the usability engineering (human factors engineering) of medical devices. The guidance is intended to support manufacturers in building a usability engineering process and to standardize the requirements for technical review. The focus is on the safe and effective use of medical devices.
Read more →
A notified body is a conformity assessment body, independent of the manufacturer and designated by a member state under Regulation (EU) 2017/745. For most risk classes it reviews the quality management system and the technical documentation and issues certificates, without which the manufacturer may not affix the CE marking.
Read more →
Packaging usability describes the extent to which a package can be used by the intended users in the intended context of use safely, effectively, and efficiently to identify, open, provide, and, where applicable, aseptically present a medical device. It considers not only the protective function of the packaging but the entire interaction between user and packaging system.
Read more →
Participant information and consent refer, in usability studies, to the notice of who processes personal data and why (Art. 13 GDPR, the EU data protection law) and to the voluntary, revocable agreement to a processing (Art. 7 GDPR). Both document the data protection concept but do not replace it.
Read more →
Perception (German: Wahrnehmung) is the first stage of human information processing: taking in signals through the senses, such as reading a display, hearing an alarm, or feeling a pressure point. What does not get through at this stage cannot be processed at any later one.
Read more →
Perception-cognition-action, PCA for short, describes basic aspects of human information processing when interacting with a product: Users must first perceive information, then interpret and process it, and finally plan and carry out an action. In usability engineering, PCA serves as a practical analysis framework.
Read more →
Post-market surveillance (PMS) is a systematic, proactive and continuous process for collecting, analyzing and assessing information about a medical device after it has been placed on the market. It serves to monitor the safety, performance and benefit-risk ratio of the product throughout its entire lifecycle. In the US, “postmarket surveillance” additionally has a narrower meaning: studies ordered by the FDA under Section 522 (see regulatory reference below).
Read more →
Posture and musculoskeletal load describe the posture a product demands of its users and the physical strain that results from it over time. Unlike a use error, this load does not show up in a single use event but only over weeks and years.
Read more →
A Pre-Submission (Pre-Sub for short) is a voluntary written request to the FDA that lets manufacturers obtain feedback on specific questions before a planned submission, optionally supplemented by a meeting. The Pre-Sub is part of the Q-Submission Program and is suitable, among other things, for agreeing on the human factors validation protocol in advance.
Read more →
A primary operating function (German: Hauptbedienfunktion) is a term that originates from earlier editions of the IEC 62366 series. It denotes a function of the product that involves user interaction and whose correct performance is relevant to the safety of the medical device.
Read more →
Pseudonymization and anonymization are two data protection terms. Pseudonymized data, for example with a participant ID, remain personal data under the GDPR (EU data protection law) because a separately stored key allows attribution. Data are anonymous only if the link to a person can no longer be established.
Read more →
The QMSR (Quality Management System Regulation) is the version of 21 CFR Part 820 that has been effective since February 2, 2026. It incorporates ISO 13485:2016 by reference. The former design controls of 820.30 are thereby replaced by the design and development requirements in Section 7.3 of the standard.
Read more →
A representative user (plural: representative users) is a person whose characteristics, abilities, experience, and limitations make them a suitable stand-in for an intended user group of a medical device; the selection must also fit the tasks and use contexts under study. In usability engineering, representative users are involved to gain insights into use and to demonstrate that the intended users can use the product safely and as intended. The term follows the intended purpose (intended use) of the device, that is, what the manufacturer specifies it is for.
Read more →
ISO 14971 describes the internationally established process for applying risk management to medical devices. The standard requires a systematic approach to identifying hazards, evaluating associated risks, implementing risk control measures and continuously monitoring residual risks across the entire product lifecycle.
Read more →
Root cause analysis (German: Ursachenanalyse; RCA) is a systematic method for identifying the underlying causes of an observed problem, use error, close call, or safety-related event. In the usability engineering of medical devices, it serves not only to document the visible use error but to identify the causes behind it at the level of the user interface, use context, user characteristics, workflows, or organizational factors.
Read more →
Characteristics related to safety, often shortened to safety-related characteristics, are qualitative or quantitative characteristics of a medical device that can affect its safety. In usability engineering, the more specific term user interface characteristics related to safety must be distinguished from them.
Read more →
Simulated use and real use studies differ in whether users operate a medical device in a usability test under simulated or under actual conditions. The FDA generally conducts human factors validation under simulated conditions. Real use comes into question when a simulation is not sufficient.
Read more →
A sterile barrier system (SBS) is the minimum packaging that prevents microorganisms from coming into contact with a sterilized medical device and maintains its sterility until the point of use, a basic concept of the ISO 11607 series. Modern requirements are no longer limited to material integrity: Whether users can actually open the system safely and present the product aseptically is itself part of the assessment.
Read more →
Substantial equivalence (SE) is the FDA's standard of review for the 510(k). The new device must have the same intended use as a legally marketed comparison device (predicate device) and, if its technology differs, must not raise different questions of safety and effectiveness.
Read more →
The summative usability evaluation is the final assessment with which a manufacturer demonstrates that the intended users can operate a medical device safely. It is carried out with the final or production-equivalent product, representative users and under realistic conditions. In regulatory usage, it corresponds to the validation of the user interface. For the US market, the FDA describes with human factors validation testing a closely related, separately defined format of evidence.
Read more →
Task analysis (German: Aufgabenanalyse) systematically breaks down the interaction between user and product into tasks, subtasks, and individual action steps. It makes visible which perception, cognition, and action demands a user must meet, and thus provides the methodological basis for identifying critical tasks and deriving realistic use scenarios.
Read more →
In the think-aloud method (German: Lautes Denken), test participants voice their thoughts, expectations and uncertainties out loud while or immediately after operating a product. The method makes the users' mental model accessible and thus provides the causes of operating problems that pure behavioral observation cannot explain.
Read more →
The UDI (Unique Device Identification) is the unique device identifier for medical devices. Made up of a device identifier (UDI-DI) and a production identifier (UDI-PI), it is placed on the label as a barcode and human readable text and stored in the UDI database in Eudamed or in the FDA's GUDID.
Read more →
The UKCA marking and MHRA (United Kingdom) refer to the market access system for medical devices in Great Britain: the MHRA regulates the market under the Medical Devices Regulations 2002, devices carry the UKCA marking or, for a limited period, the CE marking, and must be registered with the MHRA.
Read more →
Universal design (German: Design für Alle; in the European standards context also called design for all) is the design approach of laying out a product from the outset so that as many people as possible can use it without adaptation, including people with limited strength, mobility, vision, or hearing.
Read more →
Usability (German: Gebrauchstauglichkeit) describes, according to IEC 62366-1, a characteristic of the user interface that facilitates use and thereby establishes effectiveness, efficiency and user satisfaction in the intended use environment. For medical devices, usability is not only a comfort feature. Where characteristics of the user interface influence safe use, it is a safety-relevant component of product design.
Read more →
The usability engineering file (UEF; German: Gebrauchstauglichkeitsakte) is the documented entirety of the evidence, results and references from the usability engineering process of a medical device. It supports the traceability that known or foreseeable use-related hazards and hazardous situations were identified, hazard-related use scenarios were described, necessary risk control measures were implemented and their effectiveness was evaluated. The assessment of the use-related residual risk takes place in the risk management process.
Read more →
The usability engineering process (UE process) is the systematic, risk-based development process for the analysis, specification, design, verification and validation of the safe use of a medical device. In the context of IEC 62366-1, the focus is on identifying and controlling use-related risks. The process combines user research, use-related risk analysis, iterative design and evaluation into a traceably documented approach within product development.
Read more →
Use environment factors (German: Störfaktoren) are environmental and contextual factors of the use environment that can influence the interaction with a medical device, such as noise, lighting, time pressure, interruptions or wearing gloves. If tests ignore them, they systematically overestimate the actual usability.
Read more →
A use error (German: Anwendungsfehler or Nutzungsfehler) is, according to IEC 62366-1, a user action or omission of an action during the use of a medical device that leads to a result different from that intended by the manufacturer or expected by the user.
Read more →
Use requirements (also called user requirements; German: Nutzungsanforderungen) describe the properties, functions, information, and interactions that a medical device must fulfill from the perspective of the intended users so that the intended use goals can be reached safely and effectively; depending on the development goal, efficiency, comprehensibility, or ease of use can additionally be taken into account.
Read more →
A use scenario (German: Nutzungsszenario, synonym: Benutzungsszenario) describes a coherent sequence of interactions between user, medical device, and use environment to achieve a specific use goal. Within IEC 62366-1, hazard-related use scenarios are described on the basis of identified hazards and hazardous situations and serve to plan and carry out further usability evaluations; the actual risk assessment takes place in the risk management process according to ISO 14971.
Read more →
The use specification describes the essential characteristics of the intended context of use of a medical device. It summarizes the properties of patients, users, use environments and the medical field of application that are relevant for safe use.
Read more →
A use-related risk (German: nutzungsbedingtes Risiko) is a term used in practice for a risk associated with the use of a medical device. It can arise from the interplay of user, task, user interface, and use environment. Use errors or other use-related factors can be part of a foreseeable sequence of events that leads to harm via a hazardous situation. The risk results from the probability of occurrence of this harm and its severity.
Read more →
The use-related risk analysis (URRA) links user tasks, possible use errors, hazardous situations and potential harm into a continuous chain. It is the central bridge between usability engineering and risk management, the place where the design decision and the risk assessment of a medical device meet.
Read more →
The user interface (German: Benutzungsschnittstelle) comprises, according to IEC 62366-1, all means by which a user interacts with a medical device, deliberately defined broadly and extending beyond the device itself. This includes displays and controls as well as packaging, labeling, instructions for use, and training materials.
Read more →
A user profile (German: Benutzerprofil) describes the characteristics of an intended user group that are relevant to product use: qualification, experience, training level, physical and cognitive abilities, and possible limitations. It is part of the use specification and the basis for selecting representative participants in usability studies.
Read more →
User-centered design (German: Nutzerzentrierte Gestaltung) is an iterative design approach that puts the needs, abilities, and context of users at the center from the very beginning. ISO 9241-210 describes it as human-centered design. For medical devices, this approach complements the safety-oriented usability engineering process with a holistic user perspective that goes beyond mere error avoidance.
Read more →
Validation (German: Validierung) is the objective demonstration that a medical device meets the requirements of its intended purpose (intended use) and the needs of its intended users. In usability engineering, validation refers in particular to the demonstration that the final user interface can be used safely by the intended users under representative conditions of use.
Read more →
Verification (German: Verifizierung) is the objective demonstration that specified requirements have been met. In usability engineering, it can refer, for example, to testable technical requirements of the user interface specification. In the design and development process for medical devices, verification answers in particular the question: Do the design outputs meet the specified design inputs?
Read more →
Workflow analysis (German: Workflow-Analyse) examines the real work process in which the use of a medical device is embedded, including the people involved, the other devices, the information flows and the interruptions. It provides the context without which use scenarios and test setups miss reality.
Read more →